Service Manual — Matachana Autoclave

Navigate to the "Leak Test" function as per Section 4.2. The autoclave will pull a vacuum to 100 mbar and hold it for 10 minutes. The manual states the acceptable rise is <1.3 mbar/min.

The manual provides the technician passcode to access the service menu on the PLC touchscreen. (Usually a combination like Menu -> Service -> Code 1234 , but varies by model). matachana autoclave service manual

Disclaimer: This article is for informational purposes. Always follow the official documentation provided by Matachana/Getinge for your specific autoclave model. Improper servicing can lead to severe injury or equipment damage. Navigate to the "Leak Test" function as per Section 4

Always use the manual specific to your model and software version . When in doubt, call a factory-trained technician. In sterilization, the cost of a service call is insignificant compared to the cost of a failed load or a breached safety protocol. The manual provides the technician passcode to access

Refer to Section 1 – Isolate the autoclave. Close the main steam shut-off valve, shut down the electrical disconnect, and relieve chamber pressure via the manual bleed valve. Never skip this.